Micro-Flow Metering: Calibration Methods for Gear Flow Meters

Linearity in low-flow ranges is sensitive to fluid viscosity; calibrating each unit individually and providing piecewise correction factors is more reliable than using a single-point coefficient.

齿轮计量泵选型

Linearity in the microflow range is highly sensitive to fluid viscosity, temperature, and back pressure. The single-point coefficients provided by manufacturers can only serve as initial values; proper on-site calibration against the actual process fluid is required before commissioning.

I. Why Single-Point Coefficients Are Insufficient

The primary source of error in gear flow meters is internal leakage, which is directly related to differential pressure and fluid viscosity. As flow rate decreases, leakage becomes a larger proportion of total flow, so the same meter may exhibit several-fold differences in indicated error between 10% and 80% of its span.

Therefore, for small flow applications, segmented calibration across measurement ranges is recommended, covering at least three points: the lower limit, midpoint, and upper limit of typical operating flows.

II. Calibration Procedure

  • Use the actual process fluid and run under constant temperature conditions for 10 minutes; start timing once both flow rate and temperature have stabilized.

  • Collect using gravimetric methods or standard volumetric containers, with each collection period lasting no less than 3 minutes to minimize start-up/shutdown errors.

  • Repeat measurements three times at each flow point and calculate correction coefficients based on the average; if deviations are excessive, check for incomplete air venting.

  • Record temperature, viscosity, and back pressure as reference conditions for future re-calibration.

III. Segmented Correction and Re-Calibration Intervals

Enter the correction coefficients for each flow point into the control system and apply piecewise linear interpolation; this significantly improves indication accuracy in the low-flow regime. For normal operating conditions, annual re-calibration is recommended; for high-temperature, packed-bed containing, or highly corrosive service, shorten the interval to six months.